Title:Customized Titanium Mesh Based on the 3D Printed Model vs. Manual Intraoperative Bending of Titanium Mesh for Reconstructing of Orbital Bone Fracture: A Randomized Clinical Trial
Volume: 12
Issue: 3
Author(s): Shahrokh Raisian, Hamid Reza Fallahi, Kazem S. Khiabani, Mehdi Heidarizadeh*Sadaf Azdoo
Affiliation:
- Resident of Oral and Maxillofacial Surgery, Ahvaz Jundishapur University of Medical Sciences, Ahvaz,Iran
Keywords:
3D printed model, diplopia, enophthalmos, orbital fractures, surgical mesh, titanium.
Abstract: Background: This study was aimed to compare the efficacy of customized patient-specific
titanium mesh based on 3D printed model with intra-operative bending of titanium mesh for reconstructing
of orbital floor fracture.
Methods: This study was prospectively conducted on 10 patients with unilateral orbital floor fractures
caused by accident or falls. In intervention group (n=5), the CT-scan slices were used for generating
3D reconstruction of both affected and unaffected orbits. Then, a 3D printed template of mirrored unaffected
orbit was produced to mold the titanium mesh. The titanium mesh in conventional group
(n=5) was bended only manually and intraoperatively by surgeon and positioned over the bony defect.
All patients were followed-up within 1 week, 1 month and 4 months after surgery for assessing
enophthalmos, diplopia and other complications.
Results: Of 10 patients of the study only two patients in intervention group had preoperative vertical
dystopia or diplopia which had been resolved after intervention. The mean enophthalmos at baseline
and 1 week, 1 month and 4 months after surgery in control group were 3.8±0.7, 2.4±0.8, 2.4±0.8, and
2.4±0.8 mm, respectively, and in intervention group were 2.6±0.8, 0.35 ± 0.4, 0.35 ± 0.4 and 0.35 ±
0.4 mm, respectively. The mean enophthalmos did not differ significantly at baseline between two
groups, while two groups showed significant difference after surgery.
Conclusion: Finally, we concluded that the placement of patient-specific titanium implant for reconstructing
of orbital bone fracture led to better outcomes when compared to manual bending in terms of
enophthalmos and other complications.